29 Parameters

900 Touch Control

■■■■■■■■

379

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.00 - 200.00 µA

Selection

off

Default value

off

Measuring mode T:
Input range

–20.0 - 150.0 °C

Selection

off

Default value

off

Measuring mode Conc:
Input range

0.001 - 999,999.999
The unit corresponds to the one of the standards
with which the calibration of the sensor has been
carried out.

Selection

off

Default value

off

Measuring mode Cond:
Input range

0.0001 - 999.9999 mS/cm

Selection

off

Default value

off

Temperature

This parameter is not available with the commands MEAS T (temperature
measurement) and MEAS Cond (conductivity measurement).

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH and
concentration measurements.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C