3 variable list – Metrohm 900 Touch Control Manual User Manual
Page 443
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
427
29.13.3.3
Variable list
The following table contains all of the variables which can be used for cal-
culations. For variables having an index (e.g. 'EP1'), the index must be
entered manually. In the following table, this index is characterized with
the character "#".
Variable
Description
C00
Sample size
CI1, CI2
Sample identifications
The sample identifications can be used in calculations only if numerical values are entered.
DD
Duration of the entire determination
Titrants
TITER
Titer of the titrant selected in the titration command
CONC
Concentration of the titrant selected in the titration command
Titrations, measurements, calibrations
EP#
Volume/quantity of endpoint EP# (# = 1…9)
EC#
Charge at endpoint EP# (# = 1…9)
EM#
Measured value of endpoint EP# (# = 1…9)
EF#
ERC of endpoint EP# (# = 1…9)
ET#
Temperature at endpoint EP# (# = 1…9)
ED#
Time at endpoint EP# (# = 1…9)
ESI#
Recognition of endpoint EP# (# = 1…9)
Endpoint found = 1, no endpoint = 0
MIM
Initial measured value, i.e. measured value prior to the processing of the start conditions
MIT
Initial temperature, i.e. temperature prior to the processing of the start conditions
MSA
Volume for start volume
MSP
Volume for start measured value
MSS
Volume for start slope
MSV
Volume for all the start conditions
MSD
Duration of start conditions
MSM
Start measured value, i.e. measured value after the processing of the start conditions
MST
Start temperature, i.e. temperature after the processing of the start conditions
MCD
Total duration of the titration, measurement or calibration
MTM
Type of temperature measurement (Pt1000, NTC or manually)
Format = Text
MDD
Duration of effective dosing, i.e. without filling times, pauses.