beautypg.com

1 editing the user configuration, Editing the user configuration – Metrohm 900 Touch Control Manual User Manual

Page 45

background image

■■■■■■■■■■■■■■■■■■■■■■

7 System settings

900 Touch Control

■■■■■■■■

29

Operation with login function:
If you work with the login function activated, i.e. if each user must log
into the system before starting work, then only those users entered in
the list can log in. The user who is currently logged in is shown in the
main dialog.

Operation without login function:
If you work without the login function, the users whose names are
entered in the list can be selected in the main dialog or a user name
can be entered. This makes it possible to document who has operated
the titration system.

[Login options]

Define the settings for the login, Audit Trail, etc. (see Chapter 7.3.3, page
33)
.

[Create ID profile]

Create an identification profile for the selected user on a storage medium
(see Chapter 7.3.2, page 33).

[New]

Add a new user to the list (see Chapter 7.3.1, page 29).

[Delete]

Delete the selected user from the list.

NOTE

Once you have worked with the login function and password protec-
tion, users can no longer be deleted, even if the password protection is
disabled again. The status of these users must be set to inactive
(requirement of FDA guideline 21 CFR Part 11).

The last user with administrator rights cannot be deleted.

[Edit]

Edit the data of the selected user (see Chapter 7.3.1, page 29).

7.3.1

Editing the user configuration

User list: User

New / Edit