344

■■■■■■■■

900 Touch Control

The settings of the individual titration rates are listed in table 11, page
344
.

User-defined parameters
These parameters are only accessible when Titration rate = user.

Dynamics

This parameter defines the control range before the specified endpoint.
Within the control range, the bromine is generated step by step, the gen-
eration is finely controlled. The closer the endpoint, the slower the bro-
mine is generated until the rate defined under Min. rate is reached. The
larger the control range, the slower the titration. Outside the control
range, bromine is continuously generated, the rate is defined under Max.
rate
.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Input range

0.1 - 1,250.0 mV

Default value

400.0 mV

Selection

off

Max. rate

Rate at which bromine is generated outside the control range.

Input range

10.0 - 19876.0 µg/min

Default value

1000.0 µg/min

Selection

maximum

Min. rate

Rate at which bromine is generated at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.

Input range

1.0 - 999.9 µg/min

Default value

25.0 µg/min

Table 11

Default values for the predefined titration rates for BRC

Titration rate

slow

optimal

fast

Dynamics

400.0 mV

400.0 mV

400.0 mV

Max. rate

500.0 µg/min

1,000.0 µg/min

2,000.0 µg/min

Min. rate

15.0 µg/min

25.0 µg/min

50.0 µg/min