29.4 Volumetric Karl Fischer titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

326

■■■■■■■■

900 Touch Control

drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped if the endpoint has been exceeded during a
certain time period (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be stopped until the stop conditions have been
fulfilled.

Stop drift

This parameter can only be edited with Stop criterion = drift.

The titration is stopped when the endpoint and the stop drift have been
reached.

Input range

1 - 999 µL/min

Default value

20 µL/min

Delay time

This parameter can only be edited with Stop criterion = time.

When the endpoint has been reached, the specified time is allowed to
elapse after the last dosing and the titration is then stopped.

Input range

0 - 999 s

Default value

10 s

Relative stop drift

This parameter can only be edited with Stop criterion = rel. drift.

The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.

Input range

1 - 999 µL/min

Default value

10 µL/min

29.4.3

Titration parameters

Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.

Titration direction

Selection

+ | - | auto
The default value depends on the measuring mode:
Ipol = -, Upol = +