10.5 Limit values for the calibration data

■■■■■■■■■■■■■■■■■■■■■■

76

■■■■■■■■

900 Touch Control

Upper limit

pH electrodes:
Input range

–999.9 - 999.9 %

Default value

101.0 %

ISE electrodes:
Input range

–999.9 - 999.9 mV
The default value depends on the charge of the ion:
–31.0 mV (–2), –61.0 mV (–1), 61.0 mV (+1), 31.0
mV (+2).

Monitoring pH(0)

on | off (Default value: off)

This parameter is only available with pH electrodes.

If this parameter is activated, then the electrode zero point pH(0) will be
monitored.

Lower limit

Input range

–20.000 - 20.000

Default value

6.750

Upper limit

Input range

–20.000 - 20.000

Default value

7.250

Monitoring E(0)

on | off (Default value: off)

This parameter is only available with ISE electrodes.

If this parameter is activated, then the electrode zero point E(0) will be
monitored.

Lower limit

Input range

–2000.0 - 2000.0 mV

Default value

–2000.0 mV

Upper limit

Input range

–2000.0 - 2000.0 mV

Default value

2000.0 mV

Monitoring cell constant

on | off (Default value: off)

This parameter is only available for conductivity measuring cells.