2 tubing parameters, Tubing parameters – Metrohm 900 Touch Control Manual User Manual
Page 70
8.5 Exchange unit
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
900 Touch Control
Cylinder volume
The entire cylinder volume is being dosed.
Cycles
Number of rinsing cycles. We recommend carrying out at least two rinsing
cycles in order to remove all air bubbles.
Selection
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9
Default value
2
Dosing rate
Rate at which it is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488).
8.5.2
Tubing parameters
Titrant: Edit
▶ Exchange unit ▶ Tubing param.
You can enter the length and diameter of the connected tubings in the
dialog Exchange unit / Tubing parameters. The values which have
already been entered correspond to the dimensions of the supplied stand-
ard tubings.
Dosing tip
Tubing to the dosing tip (9-2).
Length
Length of the tubing.
Input range
0.0 - 999.9 cm
Default value
40.0 cm