beautypg.com

Metrohm 900 Touch Control Manual User Manual

Page 396

background image

29.9 Measurements (MEAS)

■■■■■■■■■■■■■■■■■■■■■■

380

■■■■■■■■

900 Touch Control

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Input range

0.1 - 999999.0 s

Default value

2.0 s

NOTE

The algorithm is only valid for direct measurements (MEAS Conc) if the
following conditions are fulfilled:

The sample preparation must be identical for standard solutions and
sample. In other words, if aqueous standard solutions were used for
calibration, for example, then the sample must also be produced in
aqueous medium.

The ratio sample volume / volume auxiliary solution must be identical
with the ratio volume standard solution / volume auxiliary solution
when calibrating the sensor.

Temperature compensation [Temp. compens.]
The parameters for converting the measured conductivity to a reference
temperature are defined under [Temp. compens.]. This button is avail-
able only with the MEAS Cond command (conductivity measurement).

Measuring temp.

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command).

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Reference temp.

The measured conductivity is converted to this temperature.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Selection

off

off
The temperature compensation is disabled.

Temp. compensation

Type of temperature compensation.