10 sensors – Metrohm 900 Touch Control Manual User Manual

Page 82

background image

■■■■■■■■■■■■■■■■■■■■■■

66

■■■■■■■■

900 Touch Control

10 Sensors

Main dialog: System

Sensors

This chapter describes how you can create a list of the sensors used in the
system.

Five standard sensors are defined in the sensor list: pH electrode, Metal
electrode
, Fluoride electrode, Temperature sensor and Conductiv-
ity sensor
. These sensors cannot be deleted or renamed. A maximum of
25 additional sensors can be added to these sensors.

For each sensor, the following data is displayed in the sensor list:

Designation

Type

pH: pH electrode
Metal: Metal electrode
ISE: Ion-selective electrode
Temp.: Temperature sensor
Cond.: Conductivity measuring cell

Measuring input/control device (only for intelligent sensors if they are
connected)

Intelligent sensors are also indicated by IS and are depicted in green let-
tering.

The following sensor data is stored in the list of sensors:

Name
Each sensor in the system is identified by its unambiguous name.

Calibration data (for pH electrodes, ISE electrodes and conductivity
measuring cells only)