29 Parameters

900 Touch Control

■■■■■■■■

465

Remote Box

Selection of the MSB connector the Remote Box is connected to. Every
MSB connector will always be displayed. The remote signal defined is sent
via this Remote Box.

Selection

1 | 2 | 3 | 4

Default value

1

Output signal

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.

NOTE

A line set active is not being reset automatically, not even at the end of
the determination.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.
If it is required to set pulses with other lengths, you have to define a cor-
responding template.

NOTE

We recommend masking lines that are of no interest or for which no
defined condition can be predicted with an asterisk (*).

Entry

Bit patterns containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

**************

Selection

Selection of the templates defined