29 Parameters

900 Touch Control

■■■■■■■■

453

Temperature:
Input range

–20.0 - 150.0 °C

Default value

–20.0 °C

Lower hysteresis

The lower hysteresis represents the tolerance range for the lower limit (see
Figure 27, page 455)
. When Action = Wait the dosing does not con-
tinue until the monitored quantity has once again exceeded the lower
limit incl. the value entered here.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

0.000 - 20.000

Default value

0.020

Measuring mode U:
Input range

0.0 - 2,000.0 mV

Default value

2.0 mV

Temperature:
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Upper limit

When this value is exceeded, the action defined in the following is trig-
gered.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

–20.000 - 20.000

Default value

20.000

Measuring mode U:
Input range

–2,000.0 - 2,000.0 mV

Default value

2,000.0 mV

Temperature:
Input range

–20.0 - 150.0 °C

Default value

150.0 °C