29.15 Dosing and Liquid Handling

■■■■■■■■■■■■■■■■■■■■■■

456

■■■■■■■■

900 Touch Control

commands are always sent, independent of the action defined for a limit
infringement (see "Action", page 362).

[New]

Define a new assignment.

[Delete]

Delete the selected assignment.

[Edit]

Edit the settings for the selected assignment.

Monitored quantity

Selection of the monitored quantity to which a remote signal or an
RS-232 command is to be assigned.

Selection

Measured value | Temperature | any

any
If several quantities are monitored, the remote signal or RS-232 com-
mand set is sent as soon as a limit of one of the monitored quantities is
exceeded.

Infringed limit

Indicating the limit whose infringement causes the remote signal or
RS-232 command defined in the following to be sent.

Selection

upper | lower | any | back in range

upper
The remote signal or the RS-232 command is sent when the upper
limit is exceeded.
lower
The remote signal or the RS-232 command is sent when the lower limit
is exceeded.
any
The remote signal or the RS-232 command is sent when the upper or
the lower limit is infringed.
back in range
The remote signal or the RS-232 command is sent when the value of
the monitored quantity is within the limits again (incl. hysteresis).

Interface

Selection of the Remote Box or the serial interface via which the remote
signal or the RS-232 command is sent.

Selection

Remote Box 1…4 | COM1…8