29 Parameters

900 Touch Control

■■■■■■■■

457

Output signal

This parameter can only be edited when a Remote Box is selected under
Interface.

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.

NOTE

A line set active is not being reset automatically, nor at the end of the
determination nor when the monitored quantity is in the range again.

Entering a bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.
If it is required to set pulses with other lengths, you have to define a cor-
responding template.

Entry

Bit patterns containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

**************

Selection

Selection of the templates defined

Character string

This parameter can only be edited when a serial interface (COM) is
selected under Interface.

Input of the RS-232 command as character string. All characters of the
ASCII code page can be used. Control characters (Esc, FF etc.) have to be
entered as a three-place, decimal ASCII code, starting with a slash. Each
command is automatically terminated with the ASCII characters CR and
LF.

Entry

ASCII string with a max. of 24 characters

Default value

empty