29 Parameters

900 Touch Control

■■■■■■■■

323

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.

Input range

0 - 999999 s

Default value

1 s

29.4.2

Control parameters

The control parameters for the endpoint are defined under [Control
parameters]
.

Endpoint at

Measured value for the endpoint.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode Ipol:
Input range

–2,000.0 - 2,000.0 mV

Default value

250.0 mV

Selection

off

Measuring mode Upol:
Input range

–200.00 - 200.00 µA

Default value

25.00 µA

Selection

off

Titration rate

Three predefined sets of parameters can be selected for the titration rate.

Selection

slow | optimal | fast | user

Default value

optimal