29.2 Monotonic equivalence point titrations (MET)

■■■■■■■■■■■■■■■■■■■■■■

294

■■■■■■■■

900 Touch Control

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.0 - 200.0 µA

Selection

off

Default value

off

Start slope

Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.

Measuring mode pH:
Input range

00.000 - 9,999 pH/mL

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

0 - 999 mV/mL

Selection

off

Default value

off

Measuring mode Upol:
Input range

0 - 99 µA/mL

Selection

off

Default value

off

Dosing rate

Rate to be dosed with until the start measured value or the start slope is
reached.

Input range

0.01 - 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488)
.