Metrohm 900 Touch Control Manual User Manual
Page 319
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
303
Measured value (measuring mode Upol):
Input range
–200.00 - 200.00 µA
Default value
–200.00 µA
Volume:
Input range
0.00000 - 9,999.99 mL
Default value
0.00000 mL
Upper limit
Measured value or volume for the upper limit.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Measured value (measuring mode pH):
Input range
–20.000 - 20.000
Default value
20.000
Measured value (measuring mode U, Ipol):
Input range
–2,000.0 - 2,000.0 mV
Default value
2,000.0 mV
Measured value (measuring mode Upol):
Input range
–200.00 - 200.00 µA
Default value
200.00 µA
Volume:
Input range
0.00000 - 9,999.99 mL
Default value
9,999.99 mL
EP criterion
The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.
Measuring mode pH:
Input range
0.10 - 9.99
Default value
0.50
Measuring mode U, Ipol:
Input range
1 - 999 mV
Default value
30 mV