beautypg.com

Metrohm 900 Touch Control Manual User Manual

Page 354

background image

29.5 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

338

■■■■■■■■

900 Touch Control

Conditioning

on | off (Default value: on)

If this parameter is activated, then the first time the titration is started the
working medium will be titrated to the endpoint with the specified control
parameters. The status is kept stable. The actual method run does not
begin until [

] has been pressed once more.

Start drift

If the measured drift is smaller than this value for a certain time (so-called
stabilizing time), Conditioning OK is displayed and the titration can be
started. The stabilizing time is defined in the dialog Conditioning / Con-
ditioning options
.

Input range

1 - 999 µg/min

Default value

20 µg/min

Drift correction

The endpoint amount can be corrected taking the drift value into account.
For this, the drift is multiplied with the drift correction time and this value
is then subtracted from the endpoint amount. The drift correction time is
the time interval between the end of conditioning and the end of the
determination.

Selection

auto | manual | off

Default value

auto

auto
The value of the current drift is automatically applied at the start of the
titration.
manual
If the drift is known throughout a longer period of time, this can be
entered manually.
off
No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual.

Drift for manual drift correction.

Input range

0.0 - 99.9 µg/min

Default value

0.0 µg/min

Measured value display

on | off (Default value: off)

If this parameter is activated, the currently measured value is displayed
during the conditioning.