6 cell, 7 control device, Cell – Metrohm 900 Touch Control Manual User Manual
Page 346: Control device
29.4 Volumetric Karl Fischer titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
330
■■■■■■■■
900 Touch Control
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
NOTE
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.
29.4.6
Cell
The monitoring of the reagent used is activated or deactivated under
[Cell].
Reagent monitoring
Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System
▶ Reagents.
Selection
Selection of configured reagents | off
Default value
off
off
The reagent monitoring is not active.
29.4.7
Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button is only displayed when several control devices have been
configured.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.