29.3 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

312

■■■■■■■■

900 Touch Control

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

0.001 - 20.000

Default value

2.000

Selection

off

Measuring mode U and Ipol:
Input range

0.1 - 2,000.0 mV

Default value

100.0 mV

Selection

off

Measuring mode Upol:
Input range

0.01 - 200.00 µA

Default value

10.00 µA

Selection

off

Max. rate

Rate at which dosing is carried out outside of the control range.

Input range

0.01 - 166.00 mL/min

Default value

10.00 mL/min

Selection

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488)
.

Min. rate

Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.

Input range

0.01 - 9999.00 µL/min

Default value

25.00 µL/min