29 Parameters

900 Touch Control

■■■■■■■■

317

Drift for manual drift correction.

Input range

0.0 - 99.9 µL/min

Default value

0.0 µL/min

Measured value display

on | off (Default value: off)

If this parameter is activated, the currently measured value is displayed
during the conditioning.

Conditioning options [Cond. options]

Cond. stop volume

Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [

] once again, then the titrant volume that

has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.

Input range

0.00000 - 9999.99 mL

Default value

20.0000 mL

Selection

off

Cond. stop time

Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.

Input range

1 - 999999 s

Selection

off

Default value

off

Delay reconditioning

on | off (Default value: off)

If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.

NOTE

This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working
medium.