29 Parameters
900 Touch Control
■■■■■■■■
317
Drift for manual drift correction.
Input range
0.0 - 99.9 µL/min
Default value
0.0 µL/min
Measured value display
on | off (Default value: off)
If this parameter is activated, the currently measured value is displayed
during the conditioning.
Conditioning options [Cond. options]
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [
] once again, then the titrant volume that
has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Input range
0.00000 - 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range
1 - 999999 s
Selection
off
Default value
off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
NOTE
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.