STAT titration

Problem

Cause

Remedy

The control point is
not kept stable.

The control parameters are
unsuitable.

Adjust the control parameters (see Chapter
29.7.2, page 354)
:

Define Titration rate = user and reduce
the minimum rate (Min. rate).

Define Titration rate = user and increase
the control range.

The stirring rate is too low.

Stir faster.

The arrangement of elec-
trode and buret tip is not
optimum.

Arrange the electrode and buret tip to an opti-
mum (see manual of control device).

The cylinder volume of the
exchange/dosing unit is
too large.

Use an exchange/dosing unit with a smaller
cylinder volume (yields a smaller minimum vol-
ume increment).

The electrode responds too
slowly.

Replace the electrode.

It takes an unex-
pectedly long time
for the control point
to be reached.

The control parameters are
unsuitable.

Adjust the control parameters (see Chapter
29.7.2, page 354)
:

Define Titration rate = user and reduce
the control range.

Define Titration rate = user and increase
the minimum rate (Min. rate).

Define Titration rate = user and increase
the maximum rate (Max. rate).