28 Manual control

900 Touch Control

■■■■■■■■

257

With the function [Dosing], dosing is continuous for as long as you hold
the button [Dose] pressed down.

Proceed as follows:

1 Configure the dosing function

NOTE

The dosing and filling rates should be decreased for viscous liq-
uids.

The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 32, page 488).

Enter the dosing rate.

Enter the filling rate.

2 Start dosing

Tap on [Dose].

The volume dosed is displayed. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.

3 Fill the cylinder

Tap on [Fill].

The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.

NOTE

If you exit the dialog with the fixed key [

], the dosing cylinder

will be filled automatically.

Parameter description

Dosing rate

Rate at which it is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum | dynamic

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488)
.