29.6 Bromine index determination (BRC)

■■■■■■■■■■■■■■■■■■■■■■

348

■■■■■■■■

900 Touch Control

auto
The value of the current drift is automatically applied at the start of the
titration.
manual
If the drift is known throughout a longer period of time, this can be
entered manually.
off
No drift correction takes place.

Drift value

This parameter can only be edited with Drift correction = manual.

Drift for manual drift correction.

Input range

0.0 - 999.9 µg/min

Default value

0.0 µg/min

Measured value display

on | off (Default value: off)

If this parameter is activated, the currently measured value is displayed
during the conditioning.

Conditioning options [Cond. options]

Stabilizing time

Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined in the dialog Edit command / Conditioning.

Input range

0 - 999999 s

Default value

0 s

Cond. stop time

Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.

Input range

1 - 999999 s

Selection

off

Default value

off

Delay reconditioning

on | off (Default value: off)

If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.