Metrohm 900 Touch Control Manual User Manual
Page 351
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
335
iodine is being continuously generated, the rate is defined under Max.
rate.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Input range
0.1 - 1,250.0 mV
Default value
70.0 mV
Selection
off
Max. rate
Rate at which iodine is generated outside the control range.
Input range
1.5 - 2241.0 µg/min
Selection
maximum
Default value
maximum
Min. rate
Rate at which iodine is generated at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.
Input range
0.3 - 999.9 µg/min
Default value
15.0 µg/min
Table 10
Default values of the predefined titration rates for KFC
Titration rate
slow
optimal
fast
Dynamics
120.0 mV
70.0 mV
30.0 mV
Max. rate
1,000.0 µg/min
maximum
maximum
Min. rate
0.3 µg/min
15.0 µg/min
30.0 µg/min
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 29.5.3,
page 337) always lead to a stop, even if the stop criterion has not been
reached.
Selection
drift | time | rel. drift | off
Default value
rel. drift