6 creating a subsequence (subseq), 1 loading / saving a subsequence – Metrohm 900 Touch Control Manual User Manual
Page 489

■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
473
NOTE
The sample variable can only be set on rack positions which are not
defined as special beakers.
Input range
1 - 999
Default value
1
29.17.6 Creating a subsequence (SUBSEQ)
It is useful, particularly with more complex tasks, e.g. rinsing sequences or
Liquid Handling, to combine individual command sequences into a single
subsequence. Subsequences are managed the same as method com-
mands. A maximum of 99 commands can be combined to form a single
subsequence. There is no limit to the number of subsequences within one
method.
[Load/ Save]
Loading a saved subsequence or saving the current subsequence.
[Subseq. options]
Defining various settings which correspond to the entire subsequence.
[Insert command]
Insert a new method command. It is inserted before the selected com-
mand.
NOTE
Not all commands are available for subsequences. Commands which
cannot be inserted into subsequences are disabled.
[Delete command]
Delete the selected method command.
[Edit command]
Edit the selected method command.
29.17.6.1
Loading / saving a subsequence
[Load/ Save] is used to load stored subsequences or to save new subse-
quences. This means that they are available for all methods and only have
to be created once.