11 Device manager

900 Touch Control

■■■■■■■■

93

11.5.2

Properties – Tower

List of devices: Sample Processor

Edit Tower 1/2 Properties

NOTE

In the case of a USB Sample Processor with 2 towers, Tower 1 is the
tower on the right and Tower 2 is the tower on the left, when viewed
from the front.

Maximum stroke path

Lowest permissible lift position for automatic and manual operation.

This is a safety setting. A correct specification of this value can prevent the
glass of an electrode from breaking, because a titration head can not be
moved below the position specified.

Input range

1 - 235 mm

Default value

235 mm

Min. beaker radius

Smallest permissible radius of the sample vessels for automatic and man-
ual operation.

This is a safety setting. It is intended to prevent a completely loaded, wide
titration head from being moved into a narrow sample vessel.

Input range

1.0 - 100.0 mm (Increment: 0.1)
If the lift is to be moved to the work position, a
check is made to see whether the beaker radius
defined for the attached rack (see "Beaker radius
samples", page 101)
is less than this value.

Selection

off

Default value

off