10 Sensors

900 Touch Control

■■■■■■■■

75

10.5

Limit values for the calibration data

Sensor: Edit

Limit values

You can define the following limit values in the dialog Edit sensor /
Limit values
:

Slope (pH and ion-selective electrodes)

Electrode zero point (pH and ion-selective electrodes)

Cell constant (conductivity measuring cells)

These values are monitored during the calibration. If these limits are
infringed, then a message will be displayed and you can decide whether
to accept the calibration data or not.

Monitoring slope

on | off (Default value: off)

If this parameter is activated, then the slope will be monitored.

Lower limit

pH electrodes:
Input range

–999.9 - 999.9 %

Default value

96.0 %

ISE electrodes:
Input range

–999.9 - 999.9 mV
The default value depends on the charge of the ion:
–25.0 mV (–2), –55.0 mV (–1), 55.0 mV (+1), 25.0
mV (+2).