2 messages – Metrohm 900 Touch Control Manual User Manual

Page 232

background image

24.2 Messages

■■■■■■■■■■■■■■■■■■■■■■

216

■■■■■■■■

900 Touch Control

The equivalence point criterion determined (ERC = Equivalence point Rec-
ognition Criterion) is a measure for the size of the jump in the titration
curve. The ERC determined is compared with the EP criterion defined in
the method for the evaluation of the titration curve see "Evaluation and
equivalence point criterion with DET", page 288
and
"Evaluation and
equivalence point criterion with MET", page 305)
. The ERC is thus an
important quantity when you need to adjust the parameters for the evalu-
ation of the equivalence points.

Calibration and titrant data
Results dialog: More data

View data Calibration data

Results dialog: More data

View data Titrant data

The calibration data or titrant data which was current at the time the
determination was performed is displayed in the dialogs View data / Cal-
ibration data
and View data / Titrant data.

NOTE

Recalculation:

The concentration and titer of the titrant can be modified retroactively.
The values displayed here are used for the recalculation of a loaded
determination. The values saved in the System / Titrants dialog are
not changed as a result.

24.2

Messages

Results dialog: More data

Messages

If messages occurred during a determination, they will be displayed in the
More determination data / Messages dialog. The following data for
the message is indicated under [Details]:

Message number
Unique identification number.

Time
Moment that the message occurred.

Message
Message text.