29.7 STAT titrations (STAT)

■■■■■■■■■■■■■■■■■■■■■■

364

■■■■■■■■

900 Touch Control

Upper hysteresis

Lower hysteresis

Upper limit

Lower limit

Wait

Monitored

quantity

Time

Figure 18

Action "Wait"

Communication
Under [Communication], remote signals or RS-232 commands can be
assigned to the monitored quantities. These remote signals or RS-232
commands are always sent, independent of the action defined for a limit
infringement (see "Action", page 362).

[New]

Define a new assignment.

[Delete]

Delete the selected assignment.

[Edit]

Edit the settings for the selected assignment.

Monitored quantity

Selection of the monitored quantity to which a remote signal or an
RS-232 command is to be assigned.

Selection

Measured value | Temperature | Dosing rate |
any

any
If several quantities are monitored, the remote signal or RS-232 com-
mand set is sent as soon as a limit of one of the monitored quantities is
exceeded.

Infringed limit

Indicating the limit whose infringement causes the remote signal or
RS-232 command defined in the following to be sent.

Selection

upper | lower | any | back in range