29 Parameters

900 Touch Control

■■■■■■■■

445

For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. For
the selected titrant, the validity of the titer, the working life of the titrant
and the time interval for the GLP test of the buret unit are checked at the
start of the determination.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place.

29.15.5 Monitored dosing (DOS)

With the DOS command, you can dose and at the same time monitor the
measured value and/or the temperature.

29.15.5.1

Dosing parameters

The parameters for the dosing are defined under [Dosing parameters].

Dosing criteria

Three different dosing types can be selected for the dosing. Two of the
following parameters can be specified:

Volume

Dosing rate

Dosing time

The third parameter is calculated according to the following formula:

Volume = Dosing rate · Dosing time

If continuous dosing is necessary, tandem dosing can be used (see Chap-
ter 29.7.8, page 367)
. Tandem dosing is carried out with a combination
of two dosing devices so that the second dosing device is dosing while
the first one is being filled and vice-versa.

Selection

Volume/Dosing rate | Volume/Dosing time |
Dosing rate/Dosing time

Volume/Dosing rate
Defining the volume to be dosed and and the dosing rate.
Volume/Dosing time
Defining the volume to be dosed and the dosing time.
Dosing rate/Dosing time
Defining the dosing rate and the dosing time.