8.4 Dosing unit
■■■■■■■■■■■■■■■■■■■■■■
48
■■■■■■■■
900 Touch Control
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32.1, page 488).
Dosing rate Dos. port 2
Rate used for the aspiration and ejection of the reagent via dosing port 2.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32.1, page 488).
Dosing rate Fill port
Rate used for the aspiration and ejection of the reagent via fill port.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32.1, page 488).
Dosing rate Spec.port
Rate used for the aspiration and ejection of the reagent via the special
port.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32.1, page 488).
8.4.2
Tubing parameters
Titrant: Edit
▶ Dosing unit ▶ Tubing param.
You can enter the length and diameter of the connected tubings in the
dialog Dosing unit / Tubing parameters. The values which have
already been entered correspond to the dimensions of the supplied stand-
ard tubings. In addition, the port assignment can be modified.