7 sampling, Figure 15–14. the sample-and-hold function, 1 sampling operation – Texas Instruments MSP430x1xx User Manual
Page 312: 13 × (adc12div/f
Sampling
15-22
The conversion starts with the falling edge of the sample signal SAMPCON
(see the
Sampling section and Figure 15–14). Thirteen conversion clocks
(ADC12CLK) are required to complete a conversion. The conversion time is:
t
conversion
= 13
×
(ADC12DIV/f
ADC12CLK
)
Where ADC12DIV is any integer from 1 to 8. The ADC12CLK frequency must
not exceed the maximum and minimum frequencies specified in the data
sheet. Either violation may result in inaccurate conversion results.
Note:
Availability of ADC12CLK During Conversion
Users must ensure that the clock chosen for ADC12CLK remains active
until the ADC12 can complete its operation. If the clock is removed
while the ADC12 is active, the operation can not be completed and the
end-of-conversion feedback to the program is not possible.
15.7 Sampling
The ADC12 sample-and-hold (S/H) circuitry (shown in Figure 15–14) is flex-
ible and configurable. The configuration is done by software via control bits in
the ADC12CTL0 and ADC12CTL1 registers. Configuration and operation of
the S/H circuitry is discussed in this section.
Figure 15–14. The Sample-and-Hold Function
ACLK
MCLK
SMCLK
ADC12OSC
Internal
Oscillator
ADC12CLK
S/H
Divide by
1,2,3,4,5,6,7,8
Sampling
Timer
12–bit S A R
ADC12DIV
ISSH
SHP
SHT1
SHT0
ADC12ON
ADC12SSEL
12–bit A/D converter core
analog
input
signal
Conversion CTL
MSC
SAMPCON
SHI
SHS
ENC
ADC12SC
Timer_A.OUT1
Timer_B.OUT0
Timer_B.OUT1
SYNC
VR–
VR+
Sample
Hold
and
15.7.1 Sampling Operation
The sample-and-hold circuitry samples the analog signal when the sampling
signal SAMPCON (see Figure 15–14) is high. Conversion starts immediately
with the falling edge of SAMPCON. The sample-and-hold holds the signal
value when SAMPCON is low. Conversion takes 13 ADC12CLK cycles (see
Figure 15–15).