6 meas cond - fix endpoint evaluation, 11 meas tc cond, 1 meas tc cond - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 989: Meas tc cond
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
973
can also be displayed in curves under the designation Extern 1…3 and used
in formulas as variables 'Command name.EX1…3'-
[New]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 880).
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880).
[Delete]
Deletes the external measured value selected in the table.
5.6.4.10.6
MEAS Cond - Fix endpoint evaluation
Dialog window: Method
▶ MEAS Cond ▶ Properties... ▶ Additional evalua-
tions
▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time
Default value
Measured value
Fix value
Value of the fixed endpoint.
Measured value
Range
0.0000 ... 2000.00 mS/cm
Time
Range
0.0 ... 999999.9 s
5.6.4.11
MEAS TC Cond
5.6.4.11.1
MEAS TC Cond - Overview
Dialog window: Method
▶ MEAS TC Cond ▶ Properties... ▶ MEAS TC Cond - 'Com-
mand name'
Command for the Measurement of the temperature coefficients of
the conductivity.
Devices
This command can be executed with the following devices:
Conductometer: 712, 856