4 stepping motor - stepping motor – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1226
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1210
■■■■■■■■
tiamo 2.2 (ProcessLab)
[New]
Opens the dialog window Stepping motor # in order to edit a new line
(see Chapter 5.6.11.6.4, page 1210).
[Properties]
Opens the dialog window Stepping motor # in order to edit the selected
line (see Chapter 5.6.11.6.4, page 1210).
[Delete]
Deletes the selected line.
5.6.11.6.4
STEPPING MOTOR - Stepping motor #
Dialog: Method
▶ STEPPING MOTOR ▶ Stepping motor ▶ Properties ▶ STEPPING
MOTOR - Stepping motor #
Port description
Selection of the stepping motor. The stepping motors of the selected device
are available. No port can be selected if not defined is selected in the
Device name field.
Type
Shows the type of stepping motor.
Function
Action
Action that the stepping motor is to carry out.
Selection
'empty' | Reference position | Relative position
| Absolute position | End position | Reset
Default value
'empty'
Reference position
Move to the reference position of the stepping motor.
Relative position
Move to a relative position, whereby the stepping motor moves relative
to the current position by the value defined under Distance/Angle.
Absolute position
Move to an absolute position, whereby the stepping motor moves to the
value defined under Distance/Angle.
End position
Move to the end position of the stepping motor.
Reset
A reset is carried out on the terminal the motor is connected to. While
doing so, the LEDs of the terminal will shortly flash, afterwards it could
be necessary to reference anew.