3 meas cond - measuring parameters, Measuring parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 981
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
965
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished.
5.6.4.10.3
MEAS Cond - Measuring parameters
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
Measurement frequency
712
Selection
auto | 300 Hz | 2.4 kHz
Default value
auto
auto
The more suitable of the two frequencies is selected.
Type of measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measurement
without drift control
Default value
Measurement with drift control