Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 964
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
948
■■■■■■■■
tiamo 2.2 (ProcessLab)
Parameters
The parameters for the command MEAS T/Flow are configured on the
following 4 tabs:
■
General/Hardware
Parameters for device.
■
Measuring parameters
Parameters for setting the measurement procedure.
■
Evaluation
Defining further methods for evaluating measurement curves.
■
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.
Command variables
The following command variables are generated in the method run by the
command MEAS T/Flow and can be used in formulas under the designa-
tion 'Command name.Variable designation':
Identifica-
tion
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has not ever been started
.DBL
Total duration for the processing of the command in s
.EGF
Last measured gas flow rate (measured value following
processing of the command) in mL/min
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.GMA
Maximum gas flow rate in mL/min
.GMI
Minimum gas flow rate in mL/min
.GMN
Average gas flow rate in mL/min
.IGF
Initial gas flow (measured value at the time of the start
of the command) in mL/min