Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1273
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.2 (ProcessLab)
■■■■■■■■
1257
Output signal
Entry of the bit-pattern for the output signal with exactly 14 characters.
It is possible to enter the characters
0 = Line inactive,
1 = Line active,
* = Any line status and
p = Set pulse
Selection
Bit-pattern with exactly 14 characters (0, 1, *,
p) | **************
Default value
**************
The output lines are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example: ************1* sets the output line 1 to active (= set), which
would result in a Stop command for a connected Titrino, for example.
************0* sets the line to inactive.
Note
We recommend masking the irrelevant output lines with an asterisk * so
as not to modify these line conditions.
Pulse length
Duration of the transmitted pulse.
Range
100 ... 1000 ms
Default value
200 ms
Examples for 765 Dosimat
Parameters
Binary pattern
Function
Start Dos1
*****p********
Starts Dosimat 1 (Titrino via "acti-
vate")
Start Dos2
*****p********
Starts Dosimat 2 (Titrino via "acti-
vate")
Start Dos*
*****p*p******
Starts Dosimat 1 and 2 (Titrino via
"activate")