11 det u - fix endpoint evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 507
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
491
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
EP criterion
Criterion for the recognition of equivalence points. Equivalence points
whose ERC is smaller than the set EP criterion will not be recognized.
Range
0 ... 200
Default value
5
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.3.11
DET U - Fix endpoint evaluation
Dialog window: Method
▶ DET U ▶ Properties... ▶ Additional evaluations ▶ Fix end-
point evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.