Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 573
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
557
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration rate.
Measured value acceptance
Note
Will be displayed only if the User option is selected in the Titration
rate list box.
Measured value drift
Drift for the measured value acceptance during the titration.
Titrando, 855
Range
0.1 ... 999.0 mV/min
Default value
20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Titrino
Range
0.5 ... 999.0 mV/min
Default value
20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The minimum waiting period is only important for drift-controlled measure-
ments. The measured value is not accepted until the minimum waiting time
has elapsed, even if the signal drift has already been reached. The drift con-
tinues to be checked while the waiting time is elapsing.