7 met upol - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 657
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
641
5.6.3.3.5.7
MET Upol - Additional evaluations
Tab: Method
▶ MET Upol ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value, time or volume) for the fix endpoint from the measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.3.5.11, page 648).
[Properties]
Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.3.5.11, page 648).
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.