5 tet - stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 885
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
869
Range
1 ... 200
Default value
50
Damping until
Range from the start of the measurement in which the curve is additionally
damped. This parameter allows to minimize the influence of titration arti-
facts (e.g. spikes), which occur at the beginning of the measurement and
can mistakenly be interpreted as endpoints. The damping functions as a
linear ramp, i.e. its influence is the highest at the beginning and then decrea-
ses linearly to the defined volume.
Range
0.0 ... 50.0 mL
Default value
0.2 mL
5.6.3.9.5
TET - Stop conditions
Tab: Method
▶ TET ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.
Stop volume
Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Range
0.01 ... 50.00 mL
Selection
off
Default value
off
Stop measured value
Stops when the measured value entered for a measuring point has been
reached since the start of the titration. Thereby, the direction of the meas-
urement curve (ascending or descending), which is determined after record-
ing the first 50 measuring points, is taken into account.
Range
0.0 ... 60.0 °C
Selection
off
Default value
off
Stop time
Stops when the preset time (including start conditions) has elapsed since the
start of the titration.