Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 580
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
564
■■■■■■■■
tiamo 2.2 (ProcessLab)
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is enabled, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (measured
value, time or volume) for the fix endpoint from the measuring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.3.2.11, page 571).
[Properties]
Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.3.2.11, page 571).
[Delete]
Deletes the selected line.
pK/HNP evaluation
pK/HNP evaluation
on | off (Default value: off)
If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.10.1, page 875).
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.