5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
874
■■■■■■■■
tiamo 2.2 (ProcessLab)
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.9.9
TET - Endpoint
Dialog window: Method
▶ TET ▶ Properties... ▶ Thermometric evaluation ▶ [New]/
[Properties]
▶ Endpoint
Reaction type
Selection of the reaction type according to which the curve is to be evalu-
ated.
Selection
exothermic | endothermic
Default value
exothermic
EP criterion
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
for reaction type = exothermic
Range
-999999 ... 0
Default value
-5
for reaction type = endothermic
Range
0 ... 999999
Default value
5
5.6.3.9.10
TET - Fixed endpoint evaluation
Dialog window: Method
▶ TET ▶ Properties... ▶ Additional evaluations ▶ Fixed end-
point evaluation #
Response
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fixed value
Value of the fixed endpoint.
Measured value
Range
0.0 ... 60.0 °C