Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 784
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
768
■■■■■■■■
tiamo 2.2 (ProcessLab)
time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different, very small volume increments (depending
on the volume of the exchange or dosing unit) means different switch-
off points.
Note
The Stop conditions will always stop the titration even if the stop criterion
has not been fulfilled or has been switched off.
Stop drift
Titration is stopped when the endpoint and the stop drift have been reached.
This parameter is only shown for Stop criterion = drift.
Range
1 ... 999 µL/min
Default value
20 µL/min
Relative stop drift
The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion = rel. drift.
Titrando
Range
1 ... 999 µL/min
Default value
10 µL/min
Delay time
If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is only
shown for Stop criterion = time. The selection inf. stands for infinite.
Titrando, 855
Range
0 ... 999 s
Default value
10 s
Titrino
Range
0 ... 999 s
Default value
10 s
Selection
inf.