9 kft ipol - additional measured values – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 770
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
754
■■■■■■■■
tiamo 2.2 (ProcessLab)
[Delete]
Deletes the selected line.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the minimum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated volume, time and temperature
for the maximum measured value will be interpolated from the measuring
point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
1.0 ... 2000.0 mV/mL
Default value
25.0 mV/mL
5.6.3.5.2.9
KFT Ipol - Additional measured values
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Additional measured values
Command name
Name of the command.
Input
25 characters
A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings: