5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
797
Device type
Display or selection of the device type. If a device is selected under Device
name, then this Device type field can no longer be edited, and the device
type belonging to the selected device is displayed instead.
If the option not defined is selected as Device name, any device types or
device groups which are able to execute the command can be selected,
independently of the devices in the device table.
Selection
'Device types' | Titrando
Default value
Titrando
Sensor
Sensor
Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
Polarization current at the indicator electrode.
Range
-125.0 ... 125.0 µA (Increment: 0.5)
Default value
1.0 µA
Note
When changing this parameter, make sure that the polarization current
of the indicator electrode, the endpoint and the control range are linked
closely with one another (see Chapter 5.6.3.7.4, page 800).
Electrode test
on | off (Default value: off)
If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Selection
continuous | automatic | off
Default value
automatic