5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
804
■■■■■■■■
tiamo 2.2 (ProcessLab)
Conditioning
on | off (Default value: on)
Enable/Disable the conditioning. If the conditioning is disabled, the following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series tab will then switch to COND READY.
Range
1 ... 999 µg/min
Default value
20 µg/min
Drift correction
If this parameter is set to automatic or manual the drift correction ( = Drift
value × Time for drift correction) caused by empty consumption during
titration is deducted from the total consumption. The Time for drift cor-
rection is the period between end of conditioning and end of titration.
Selection
automatic | manual | off
Default value
off
automatic
The drift lastly determined during conditioning will be used as drift value.
manual
The drift value defined under Drift value will be used.
off
No drift correction takes place.
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 999.9 µg/min
Default value
0.0 µg/min
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Range
1 ... 999999 s
Selection
off
Default value
off