4 properties - glp – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1397
■■■■■■■■■■■■■■■■■■■■■■
7 Devices
tiamo 2.2 (ProcessLab)
■■■■■■■■
1381
Note
These parameters are only editable for devices with status not ok (power
supply and/or RS connection interrupted).
7.3.4
Properties - GLP
Tab: Configuration
▶ Device ▶ [Edit] ▶ Properties... ▶ 'Properties - 'Device type' -
'Device name'
▶ GLP
GLP test date
Date of the last GLP test. This date can be selected by clicking on
in the
dialog window Select date (see Chapter 2.5.1, page 83).
Comment GLP test
Comment on GLP test.
Input
1000 characters
Monitoring of GLP validity
Monitoring of GLP validity
on | off (Default value: off)
If this option is switched on then the time interval for the GLP test will be
monitored.
Monitoring can only be switched on when a date has been entered in the
field GLP test date.
GLP test interval
Time interval to next GLP test. If a value is entered here then the date in the
field Next GLP test will be adapted automatically.
Only editable when the monitoring is activated.
Range
1 ... 999 Days
Default value
999 Days
Next GLP test
Date on which the next GLP test is to be carried out. The date can be selected
by clicking on the symbol
in the dialog window Select date (see Chapter
2.5.1, page 83). After the date has been entered, the field GLP test inter-
val will be adapted automatically.