2 calc - calculation overview – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1158
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1142
■■■■■■■■
tiamo 2.2 (ProcessLab)
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
never been started
.FIN
Command status; 1 = Command has ended at least
once, 0, invalid (variable not available) = Command has
never ended
Result variables
The following result variables are defined in the CALC command and can
be used in formulas under the designation 'RS.ResultName.Variable
designation':
Identifica-
tion
Description
.VAL
Result value, optional, i.e. 'RS.RS01' = 'RS.RS01.VAL'
.ASD
Absolute standard deviation for the result
Max.
Maximum value of the result
min.
Minimum value of the result
.MNV
Mean value of the result
.NSR
Statistics actual counter for the result
.NST
Statistics setpoint counter for the result
.OVF
Limit value violation for result; 1 = limit exceeded, 0 =
limit not exceeded
.RSD
Relative standard deviation for the result
.STS
Statistics status for the result; 1 = Statistics on, 0 = Sta-
tistics off
.UNI
Result unit (Text)
5.6.8.2.2
CALC - Calculation overview
Dialog window: Method
▶ CALC ▶ Properties... ▶ CALC 'Command name'
Command name
Name of the command.
Input
25 characters
Moves the selected result up (changes the sequence).