5 calc - result properties, 1 result - definition, Definition – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1162
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1146
■■■■■■■■
tiamo 2.2 (ProcessLab)
Enter constant
All of the constants of the formula template will be listed.
'Constant'
Entering the constants.
Range
-1.0 E+99 ... 1.0E+99 (max. 15 places)
Default value
'empty'
Opens molar mass calculator for entering molar masses (see Chapter
2.4.5, page 83).
5.6.8.2.5
CALC - Result properties
5.6.8.2.5.1
Result - Definition
Tab: Method
▶ CALC ▶ Properties... ▶ [New] ▶ [Continue] ▶ Result - Definition
The result calculation is defined on this tab.
Result name
Name of the result. This name will be shown in the result view as well as in
the report. The result name can be used in further calculations as variable
RS.Result name.VAL.
Input
50 characters
Default value
Res01 … Res025
Formula
Display of the calculation formula of the result. The formula editor to enter
the calculation formula is opened with a click on
or by double-clicking
in the window in which the formula can be entered (see Chapter 2.4, page
25). The result of the calculation formula defines the result type (Number,
Text or Date/time).
Input
1000 characters
Unit
Unit of the result for the output (text only). The unit can be used in further
calculations as variable 'RS.Result name.UNI'.
Input
16 characters
Selection
empty | ppm | % | g/L | mg/L | mg/mL | mg/100
g | mol/L | mmol/L | mL | µL | g | mg | µg | °C | s |
mV | µA | S/cm | mS/cm | µS/cm | µL/min
Default value
empty